Bowie Dick Test Pack (TP01)

The TP01 Bowie-Dick Test Pack is a Type 2 indicator, specifically designed for the daily monitoring of steam autoclaves equipped with vacuum pumps and pre-vacuum cycles. It is a critical tool for detecting leaks, inadequate steam penetration, air pockets, and vacuum pump failures.

Performance and Diagnostics:

  • Precision Evaluation: Its design creates a controlled resistance to steam to simulate a real load, evaluating whether the air removal system is functioning correctly.

  • Interpretation of Results: A uniform and complete color change confirms that air has been evacuated and steam has penetrated effectively. If areas without color change or partial discolorations appear, it indicates residual air or system failures.

  • Traceability: The sensitive ink test sheet provides a permanent color change, allowing it to be kept as a physical record of the cycle validation.

Technical Specifications:

  • Exposure Conditions:

    • $134°C$ for 3.5 minutes ($±30$ sec).

    • $121°C$ for 15 minutes.

  • Regulation and Quality: Designed under the ISO 11140-1, 4 standard. Processes certified by ISO 13485 and Good Manufacturing Practices (GMP).

  • Shelf Life: 60 months. Non-reusable product.